Reliability of goniometric measurements in children with cerebral palsy: A comparative analysis of universal goniometer and electronic inclinometer. A pilot study

by: Pablo Herrero, Patricia Carrera, Elena Garcia, Eva G. Trullen, Barbara O. Blazquez
BMC Musculoskeletal Disorders, Vol. 12, No. 1. (10 July 2011), 155, doi:10.1186/1471-2474-12-155  Key: citeulike:9529327

Formatted Citation


Show HTML

Likes (beta)

This copy of the article hasn't been liked by anyone yet.

View FullText article


Abstract

BACKGROUND:Even though technological progress has provided us with more and more sophisticated equipment for making goniometric measurements, the most commonly used clinical tools are still the universal goniometer and, to a lesser extent, the inclinometer. There is, however, no published study so far that uses an inclinometer for measurements in children with cerebral palsy (CP). The objective of this study was two-fold: to independently assess the intra and inter-examiner reliability for measuring the hip abduction range of motion in children with CP using two different instruments, the universal two-axis goniometer and electronic inclinometer. A pool of 5 examiners with different levels of experience as paediatric physiotherapists participated. The study did not compare both instruments because the measurement procedure and the hip position were different for each.METHODS:A prospective, observational study of goniometery was carried out with 14 lower extremities of 7 children with spastic CP. The inclinometer study was carried out with 8 lower extremities of 4 children with spastic CP. This study was divided into two independent parts: a study of the reliability of the hip abduction range of motion measured with a universal goniometer (hip at 0degrees flexion) and with an electronic inclinometer (hip at 90degrees flexion). The Intraclass Correlation Coefficient (ICC) was calculated to analyse intra and inter-examiner agreement for each instrument.RESULTS:For the goniometer, the intra-examiner reliability was excellent (>0.80), while the inter-examiner reliability was low (0.375 and 0.475). For the inclinometer, both the intra-examiner (0.850-0.975) and inter-examiner reliability were excellent (0.965 and 0.979).CONCLUSIONS:The inter-examiner reliability for goniometric measurement of hip abduction in children with CP was low, in keeping with other results found in previous publications. The inclinometer has proved to be a highly reliable tool for measuring the hip abduction range of motion in children with CP, which opens up new possibilities in this field, despite having some measurement limitations.


DFellNeuroPT's tags for this article

Citations (CiTO)

No CiTO relationships defined

X There are no reviews yet

X Find related articles with these CiteULike tags

X Posting History


X Export records

Privacy Statement | Terms & Conditions
CiteULike organises scholarly (or academic) papers or literature and provides bibliographic (which means it makes bibliographies) for universities and higher education establishments. It helps undergraduates and postgraduates. People studying for PhDs or in postdoctoral (postdoc) positions. The service is similar in scope to EndNote or RefWorks or any other reference manager like BibTeX, but it is a social bookmarking service for scientists and humanities researchers.