CiteULike | ![]() |
rairighd's CiteULike | ![]() |
![]() |
|
![]() |
Register | ![]() |
Log in | ![]() |
Real-time high-sensitivity impedance measurement interface for tethered BLM biosensor arrays |
Reviews
[Write a review of this article]
Notes for this articleInterrogate a tBLM using a matched impedance to cancel the baseline.
Find related articles from these CiteULike users
Find related articles with these CiteULike tags
Posting History
AbstractThis paper presents a switched-capacitor (SC) current integrator circuit for impedance measurement of tethered bilayer lipid membrane (tBLM) biosensors. The circuit comprises a small number of high performance components enabling enhanced experimental flexibility and reliability. The sensitivity is improved significantly by suppressing the output offset through pseudo-differential operation, using R-C components for the reference impedance. The sensing and reference electrodes are excited with low-amplitude differential voltage pulses and the current response to membrane resistance (R<inf>M</inf>) change of the tBLM biosensor is converted to voltage by a precision, low-noise SC integrator available as a single-package IC. Tests with both electrical models and actual biosensors demonstrated that the proposed circuit operates with high sensitivity and can be used in single chip versions for low-cost and high-sensitive tBLM biosensor arrays, featuring multiple electrode sites.
BibTeX record
RIS record