A Programmable 0.18-μm CMOS Electrochemical Sensor Microarray for Biomolecular Detectionby: A. Hassibi, T. H. Lee
|
Reviews
[Write a review of this article]
Notes for this articleThey claim impedance measurement, but actually only provide a reconfigurable amplifier. Array size: 5x10; Scan rate: 0.5 samples/s; Imped. Freq. range: 50kHz - 10,000 kHz
See also: "A programmable electrochemical biosensor array in 0.18 μm standard CMOS" from ISCC
Find related articles from these CiteULike users
Find related articles with these CiteULike tags
Posting History
AbstractA configurable electrochemical sensor microarray system-on-a-chip fabricated in a standard digital 0.18-mum complementary metal-oxide-semiconductor (CMOS) process is presented. Each pixel within this 5times10 array occupies a 160 mumtimes120 mum area and contains a differential electrochemical transducer with a programmable sensor. The sensor has a digitally configurable topology capable of performing different electroanalytical measurements for a variety of affinity-based biomolecular sensing applications. The main modes of operation for this system are impedance spectroscopy, voltammetry, potentiometry, and field-effect sensing