A depth of interaction detector for PET with GSO crystals doped with different amounts of Ceby: N. Inadama, H. Murayama, T. Omura, T. Yamashita, S. Yamamoto, H. Ishibashi, H. Kawai, K. Omi, T. Umehara, T. Kasahara
Nuclear Science, IEEE Transactions on In Nuclear Science, IEEE Transactions on, Vol. 49, No. 3. (2002), pp. 629-633.
|
Reviews
[Write a review of this article]
Find related articles from these CiteULike users
Find related articles with these CiteULike tags
Posting History
AbstractA new method for a four-stage depth of interaction (DOI) detector is proposed. The four-stage DOI detector is constructed with two kinds of Gd/sub 2/SiO/sub 5/:Ce (GSO) crystals doped with different amounts of Ce, 0.5 mol% and 1.5 mol%. The amount of Ce in GSO determines the scintillation decay time constant, and it is 60 ns for the 0.5 mol% GSO and 35 ns for 1.5 mol% GSO. This difference led to the introduction of pulse-shape discrimination, which would distinguish between two kinds of event data from respective GSOs and sort them into two groups. By independently applying Anger-type position arithmetic to the data of each group, two two-dimensional (2-D) histograms are obtained. The crystal of interaction can be identified on these histograms in which only 0.5 mol% or 1.5 mol% GSO crystal elements are expressed. To evaluate this method, we constructed the four-stage DOI detector by alternately stacking 1.5 mol% GSO crystal stages and 0.5 mol% stages. The result of a scanning measurement with a /sup 137/Cs gamma-ray beam proved that the DOI detector had enough accuracy in crystal identification.