CiteULike is a free online bibliography manager. Register and you can start organising your references online.

Depth of interaction measurement in gamma ray imaging detectors with continuous scintillation crystals Export

Nuclear Science Symposium Conference Record, 2003 IEEE In Nuclear Science Symposium Conference Record, 2003 IEEE, Vol. 3 (2003), pp. 2169-2173 Vol.3.

Citation Format

[Posts]

View FullText article


rbmoura's tags for this article

crystal detector doi gamma-ray imaging scintillator

X Reviews [Write a review of this article]

X Find related articles from these CiteULike users

X Find related articles with these CiteULike tags

X Posting History

X Abstract

A design for an inexpensive depth of interaction (DOI) detector for gamma rays, suitable for nuclear medical applications, especially positron emission tomography (PET), has been developed, studied by simulations and tested experimentally. The detector consists of a continuous LSO-scintillator of dimensions 42/spl times/42/spl times/10 mm/sup 3/ and a new compact large-area (49/spl times/49 mm/sup 2/) position sensitive photo-multiplier (PSPMT) H8500 from Hamamatsu. Since a continuous crystal is used, the scintillation light distribution is not destroyed and its first 3 moments can be used to determine the energy (0th moment), the centroids along the x- and y-direction (1st moments) and the depth of interaction (DOI), which is strongly correlated to the distribution's width and thus its standard deviation (2nd moment). The simultaneous computation of these moments allows a three-dimensional reconstruction of the position of interaction of the /spl gamma/-rays within the scintillating crystal and will be realized by a modified position sensitive proportional (PSP) resistor network. No additional photo detectors or scintillating crystals are needed. According to previous Monte Carlo simulations which estimated the influence of Compton scattering for 511 keV /spl gamma/-rays, the transport of the scintillation light within the detector assembly and also the behavior of the modified PSP resistor network, we expect a spatial resolution of /spl lsim/ 2 mm and a DOI resolution of /spl ap/ 5 mm. The first experimental results presented here yield an intrinsic spatial resolution of /spl lsim/ 1.8 mm and 2.6 mm for the x- and y-direction respectively and a DOI resolution /spl lsim/ 1 cm. Further we measured an energy resolution of 12%-18%.


X BibTeX record

X RIS record


Privacy Statement | Terms & Conditions
CiteULike organises scholarly (or academic) papers or literature and provides bibliographic (which means it makes bibliographies) for universities and higher education establishments. It helps undergraduates and postgraduates. People studying for PhDs or in postdoctoral (postdoc) positions. The service is similar in scope to EndNote or RefWorks or any other reference manager like BibTeX, but it is a social bookmarking service for scientists and humanities researchers.