Analysis of quality control data of eight modern radiotherapy linear accelerators: the short- and long-term behaviours of the outputs and the reproducibility of quality control measurements. Export

Phys Med Biol, Vol. 51, No. 14. (Jul 2006), pp. 3581-3592.

Citation Format

[Posts]

View FullText article


tertel's tags for this article

accelerators calibration computer-assisted control dosage factors humans imaging models of particle phantoms photons planning quality radiometry radiotherapy reproducibility results risk statistical time

X Reviews [Write a review of this article]

X Find related articles from these CiteULike users

X Find related articles with these CiteULike tags

X Posting History

X Abstract

Quality control (QC) data of radiotherapy linear accelerators, collected by Helsinki University Central Hospital between the years 2000 and 2004, were analysed. The goal was to provide information for the evaluation and elaboration of QC of accelerator outputs and to propose a method for QC data analysis. Short- and long-term drifts in outputs were quantified by fitting empirical mathematical models to the QC measurements. Normally, long-term drifts were well (< or =1%) modelled by either a straight line or a single-exponential function. A drift of 2% occurred in 18 +/- 12 months. The shortest drift times of only 2-3 months were observed for some new accelerators just after the commissioning but they stabilized during the first 2-3 years. The short-term reproducibility and the long-term stability of local constancy checks, carried out with a sealed plane parallel ion chamber, were also estimated by fitting empirical models to the QC measurements. The reproducibility was 0.2-0.5% depending on the positioning practice of a device. Long-term instabilities of about 0.3%/month were observed for some checking devices. The reproducibility of local absorbed dose measurements was estimated to be about 0.5%. The proposed empirical model fitting of QC data facilitates the recognition of erroneous QC measurements and abnormal output behaviour, caused by malfunctions, offering a tool to improve dose control.


X BibTeX record

X RIS record


Privacy Statement | Terms & Conditions
CiteULike organises scholarly (or academic) papers or literature and provides bibliographic (which means it makes bibliographies) for universities and higher education establishments. It helps undergraduates and postgraduates. People studying for PhDs or in postdoctoral (postdoc) positions. The service is similar in scope to EndNote or RefWorks or any other reference manager like BibTeX, but it is a social bookmarking service for scientists and humanities researchers.