Evaluation of the gamma dose distribution comparison method. Export

Med Phys, Vol. 30, No. 9. (Sep 2003), pp. 2455-2464.

Citation Format

[Posts]

View FullText article


tertel's tags for this article

algorithms and computer-assisted conformal control dosage gamma humans of planning processes quality radiometry radiotherapy rays reproducibility results sensitivity software specificity stochastic

X Reviews [Write a review of this article]

X Find related articles from these CiteULike users

X Find related articles with these CiteULike tags

X Posting History

X Abstract

The gamma tool was developed to quantitatively compare dose distributions, either measured or calculated. Before computing gamma, the dose and distance scales of the two distributions, referred to as evaluated and reference, are renormalized by dose and distance criteria, respectively. The renormalization allows the dose distribution comparison to be conducted simultaneously along dose and distance axes. The gamma quantity, calculated independently for each reference point, is the minimum distance in the renormalized multidimensional space between the evaluated distribution and the reference point. The gamma quantity degenerates to the dose-difference and distance-to-agreement tests in shallow and very steep dose gradient regions, respectively. Since being introduced, the gamma quantity has been used by investigators to evaluate dose calculation algorithms, and compare dosimetry measurements. This manuscript examines the gamma distribution behavior in two dimensions and evaluates the gamma distribution in the presence of data noise. Noise in the evaluated distribution causes the gamma distribution to be underestimated relative to the no-noise, condition. Noise in the reference distribution adds noise in the gamma distribution in proportion to the normalized dose noise. In typical clinical use, the fraction of points that exceed 3% and 3 mm can be extensive, so we typically use 5% and 2-3 mm in clinical evaluations. For clinical cases, the calculation time is typically 5 minutes for a 1 x 1 mm2 interpolated resolution on an 800 MHz Pentium 4 for a 14.1 x 15.2 cm2 radiographic film.


X BibTeX record

X RIS record


Privacy Statement | Terms & Conditions
CiteULike organises scholarly (or academic) papers or literature and provides bibliographic (which means it makes bibliographies) for universities and higher education establishments. It helps undergraduates and postgraduates. People studying for PhDs or in postdoctoral (postdoc) positions. The service is similar in scope to EndNote or RefWorks or any other reference manager like BibTeX, but it is a social bookmarking service for scientists and humanities researchers.