Test liquids for quantitative MRI measurements of self-diffusion coefficient in vivo Export

Magnetic Resonance in Medicine, Vol. 43, No. 3. (2000), pp. 368-374.

Citation Format

[Posts]

View FullText article


xtizon's tags for this article

calibration dw-mri mri phantom quality

X Reviews [Write a review of this article]

X Find related articles from these CiteULike users

X Find related articles with these CiteULike tags

X Posting History

X Abstract

A range of liquids suitable as quality control test objects for measuring the accuracy of clinical MRI diffusion sequences (both apparent diffusion coefficient and tensor) has been identified and characterized. The self-diffusion coefficients for 15 liquids (3 cyclic alkanes: cyclohexane to cyclooctane, 9 n-alkanes: n-octane to n-hexadecane, and 3 n-alcohols: ethanol to 1-propanol) were measured at 15-30°C using an NMR spectrometer. Values at 22°C range from 0.36 to 2.2 10-9 m2s-1. Typical 95% confidence limits are ±2%. Temperature coefficients are 1.7-3.2 %/°C. T1 and T2 values at 1.5 T and proton density are given. n-tridecane has a diffusion coefficient close to that of normal white matter. The longer n-alkanes may be useful T2 standards. Measurements from a spin-echo MRI sequence agreed to within 2%. Magn Reson Med 43:368-374, 2000. © 2000 Wiley-Liss, Inc.


X BibTeX record

X RIS record


Privacy Statement | Terms & Conditions
CiteULike organises scholarly (or academic) papers or literature and provides bibliographic (which means it makes bibliographies) for universities and higher education establishments. It helps undergraduates and postgraduates. People studying for PhDs or in postdoctoral (postdoc) positions. The service is similar in scope to EndNote or RefWorks or any other reference manager like BibTeX, but it is a social bookmarking service for scientists and humanities researchers.